home > Product

Isotope5 CN Isotope Ratio Mass Spectrometer
Share This Page :
       The Isotope5 CN Isotope ratio mass spectrometer is a small mass spectrum analyzer produced by the British company Compact Science Systems introduced by PRI-ECO. Isotope5 breaks many limitations of the use of mass spectrometers, is small and light, and can be placed on a laboratory desktop. The performance of isotope is stable and reliable, and high quality isotope data can be quickly measured. Simple and fast operation, no need for professional and technical personnel to run; The operation cost is economical, and only needs less helium, standard gas and electricity.
       The Isotope5 CN mass spectrometer and the Element5 analyzer can be combined to form a full-performance EA-IRMS for elemental and isotope analysis of solid and liquid samples. It can be connected with gas chromatography to form a stable and reliable GC-IRMS for material content and isotope analysis of gaseous samples. It can also be connected with liquid chromatography to form fast and efficient LC-IRMS for the substance content and isotope analysis of liquid samples. Among them, EA-IRMS is equipped with Element5 analyzer, which can measure the CN content of the sample, and provide high-purity CO2 and N2 of the Isotope5 for the analysis of CN isotope. Element5 can also be used independently to measure the CNHS or O element content of a sample.
       The Isotope5 CN mass spectrometer allows rapid detection of samples containing CN in application scenarios such as material cycling (e.g., CN cycling studies), food authenticity (e.g., honey adulteration), and disease diagnosis (e.g., Helicobacter pylori identification). It can help people make new discoveries in the fields of climate research, biochemistry, forensic medicine, food analysis, oil and gas exploration, and soil science.

Technical principles
       Isotope measurement based on molecular (atomic) mass number

Key features
  • High sensitivity, accuracy and precision
  • Low gas consumption, standby mode
  • Simple and efficient installation, maintenance-free operation
  • Small size, good performance, simple operation
Performance indicators
Isotope5 CN technical parameters
Internal reproducibility of CO2 isotope 13C ±0.10 ‰ (natural abundance, 1 SD) @ CO2 standard gas injection was repeated 12 times at a mass 44 ion beam and an intensity of 20 nA amp
Internal repeatability of N2 standard gas 15N ±0.15 ‰ (natural abundance, 1 SD) @ Nitrogen standard gas was injected 12 times with a mass of 28 ion beam and an intensity of 20 nA amp
Urea standard sample 13C reproducibility ±0.2 ‰ (natural abundance, 1 SD) @ 5 urea standard samples containing 100 ug of carbon were repeated
Urea standard sample 15N reproducibility ±0.3 ‰ (natural abundance, 1 SD) @ 5 urea standard samples containing 100 ug of carbon were repeated
Sample analysis time 4 ~ 5 minutes for a sample, depending on the element analyzer or the entire system
Resolution Mass 29 >75
System Parameters
Ion beam detection CNHS Triple Faraday collector Mass Analyzer Construction 14 cm RAD, 90 degrees
Mass Analyzer magnet

Permanent high temperature  stabilizing magnet

resolution Center collector 80
Vacuum unit Built-in vacuum pump Low power consumption Typical power 240 W
Size Height 47 cm, length 70 cm, width 30 cm Weight 45 kg
Data system Includes instrument diagnostic procedures, preparation system control, analytical data acquisition and result display Software Full featured software package: for mass spectrometer control and isotope ratio analysis

Technical parameters of Element5 820
Measuring range C: 0.002-20 mg;  H: 0.002-5 mg; N: 0.002-20 mg; S: 0.002-6 mg;  O: 0.002-2 mg Analysis time CN: 5 min;  CHN: 8 min; CHNS: 10/25 min; O: 4 min
Accuracy <0.2 % (standard, purity >99.9 %) Reactants Dual furnace system
Precision <0.1 % (standard, purity >99.9 %) Display Touch screen display
Sampler Pneumatic automatic sampler: 147 bit; Electric automatic sampler: 32,50,100 bit; Manual sampler
System parameters
Dimensions 810 x 500 x 370 mm Weight 68 kg
Power supply 230 V, 50/60 Hz Power consumption 5 A, 1100 W
Gas requirements Helium (99.999 %), 3-5 bar; Oxygen (99.999 %), 3-5 bar; Air (oil-free compressed air)
Analytical conditions
Carrier Gas Helium Leak detection Automatic leak detection
Reactor temperature Left furnace: maximum 1100 ℃; Right furnace: maximum 1100 ℃ Separator furnace temperature Max. 110 ℃
Oxygen requirements Automatically calculated   from the oxygen gauge Flow regulation Electronic flow regulation
Gas separation 0.8-4 m GC columns Detector High sensitivity TCD
Software EAS Clarity Calibrate Linear, conic, cubic
Sample size 0.1-500 mg (depending on sample nature), maximum soil sample injection up to 1000 mg Sample Type Solid and liquid
Bag sample High purity tin cup or silver cup Optional accessories Scales, consumables

Manufacturer: PRI-ECO, China